Status: concept design. Nothing here has been built, verified or cleared. Every performance number is a design target taken from a named precedent, and the business plan says what happens if a target is missed.
In plain English: Instead of one magic machine, the Station combines four kinds of measurement that already work in other products. Each gets its own small share of the blood. The Station checks every sample for problems and refuses to report a number it cannot trust.
The Thimble Station is a counter-top analyzer for pharmacies and urgent care clinics. It runs about 40 assays plus a 19-parameter complete blood count from a capillary sample of roughly 300 microliters, with all in-store results in about 15 minutes, and sends anything it cannot run well to a central "hub" laboratory the same day. It does this with four kinds of measurement engine that have each been cleared or CE-marked by other companies, arranged so that no single engine has to do everything.
1. Requirements
Requirement
Target
Precedent that shows it is possible
Sample
About 300 µL capillary blood (upper arm or fingertip), or venous
BD MiniDraw: 16 to 18 drops for 16 analytes. Tasso+: 200 to 600 µL from the upper arm.
In-store menu
About 40 assays plus a 19-parameter CBC
Piccolo: 31 chemistry analytes. Boditech: dozens of immunoassays. HemoScreen and OLO: 5-part CBC.
Time to result
15 minutes or less for everything run in store
Piccolo 12 min, HemoScreen 6 min, ichroma 3 to 15 min.
Operator
Pharmacy technician after about 2 hours of training; CLIA-waived modules by phase D
11 waived Piccolo discs, Sysmex XW-100 (waived CBC), Afinion and DCA (waived HbA1c), i-STAT CHEM8+.
Accuracy
Inside the CLIA acceptance limits that took effect in 2025, with a sigma metric of 3 or better
A cleared predicate exists for every analyte.
Throughput
3 to 4 patients per hour per Station
One chemistry disc per 12 minutes.
Footprint
Under 0.2 m² of counter, under 15 kg
Piccolo is about 5 kg; Truvian is desktop-computer sized.
Connectivity
HL7 and FHIR results, remote QC, works offline
Piccolo EMR export; Athelas cloud results.
2. Design principles
Split the sample by assay class at the moment of collection. EDTA for cells, lithium heparin for chemistry and immunoassays, serum only when the hub lab needs it.
No new physics. Four engine types, each with a cleared or CE-marked precedent.
Native volumes. Each consumable meters its own sample. No dilution beyond the validated protocol, no robot pipettor.
Measure pre-analytical quality on every sample. Fill, age, clots, hemolysis, lipemia, icterus.
Suppress, never massage. A result the Station cannot stand behind is not reported. The patient is offered a venous draw or the hub lab.
Everything traceable. Lot, calibration, QC status, operator, sample age and interference indices travel with every result.
The long tail goes to the hub. The Station does the high-volume routine menu; specialty tests ride a courier.
3. System overview
System overview. One capillary draw is split into two tubes and run on four bays in about 15 minutes; results and QC go to the cloud, long-tail tests go to the hub lab.Plain-text version
PATIENT
| upper-arm vacuum collector (300-500 µL) or fingertip by trained staff (16-18 drops)
v
+----------------------------------- THIMBLE DRAW KIT -----------------------------------+
| first drop wiped -> [ EDTA microtube ~120 µL ] [ Li-heparin microtube ~250 µL ] |
| [ SST microtube ~300 µL, only when hub tests are ordered ] |
+----------------------------------------------------------------------------------------+
| EDTA | Li-heparin | SST (optional)
v v v
+---------------------- THIMBLE STATION (counter-top) ----------------------+ +--- HUB LAB ---+
| BAY 1 HEME CARD BAY 2 CHEM DISC BAY 3 IMMUNO PACK | | CAP / CLIA |
| imaging CBC + HbA1c centrifugal photometry time-resolved | | courier, |
| 20 µL, ~5 min 100 µL, ~12 min fluorescence strips | | same day, |
| 15 µL each, 10-15 min | | long-tail |
| BAY 4 (urgent-care option) EC CARD: Na K Cl iCa Glu Lac Crea BUN Hct | | menu |
| | +-------+-------+
| pre-analytics: fill | sample age | clots | hemolysis, lipemia, icterus | |
| CM5 + NPU compute | real-time MCU | NFC lot data | barcode | touchscreen | |
+-------------------------------------+-------------------------------------+ |
| encrypted results + QC records |
v v
+---------------------------- THIMBLE CLOUD ----------------------------+
| fleet QC (Levey-Jennings, patient moving averages) | remote lockout |
| lab director review | HL7 / FHIR to EHRs | patient app | audit trail |
+-----------------------------------------------------------------------+
4. Collection and the volume budget
The Draw kit is built from components that are already FDA-cleared: a contact-activated safety lancet or an upper-arm vacuum collector (Tasso+ and TAP Micro Select are cleared), and capillary microtubes with dry EDTA and lithium heparin (BD MiniDraw, BD Microtainer or Sarstedt Microvette class). Theranos's nanotainer was cited by FDA as an uncleared Class II device; the lesson is to buy or clear the collection hardware, never to ship it on an exemption.
Collection protocol (CLSI GP42 order of draw for capillary samples): warm the site, let the alcohol dry for 30 seconds, puncture, wipe the first drop, fill EDTA first (to limit platelet clumping), then heparin, then serum if ordered, without squeezing. Invert each tube ten times. Scan the kit barcode, which timestamps the draw.
Typical pharmacy panel: CBC and HbA1c, metabolic and lipid disc, three immunoassay strips. The Heme Card figure includes 5 µL of dead volume.
Show the data as a table
Tube
Use
Volume (µL)
EDTA tube, ~120 µL
Heme Card
25
EDTA tube, ~120 µL
Reserve
95
Li-heparin tube, ~250 µL
Chem Disc
100
Li-heparin tube, ~250 µL
Immuno, 3 strips
45
Li-heparin tube, ~250 µL
Reserve
105
If both chemistry discs and several strips are ordered, collect about 350 µL of heparin blood instead of 250 µL.
For comparison: Theranos promised more than 200 tests from "a few drops." Truvian's cleared instrument uses a full venous tube. Babson and BD use 16 to 18 drops for their capillary menu and send them to a lab.
5. Engine 1: Heme Card (imaging hematology plus HbA1c)
Precedents: Sight OLO (fingerstick CBC, 510(k) 2019), PixCell HemoScreen (20 µL, direct fingerstick 510(k) 2023), HemoCue WBC DIFF (10 µL, 5-part differential by imaging), HORIBA Microsemi (CBC from 10 µL), Athelas (white count and neutrophils by computer vision), Abbott Afinion (HbA1c by boronate affinity from 1.5 µL).
Heme Card: four channels from one 20 µL inlet, imaged from below.Plain-text version
HEME CARD (75 x 25 x 3 mm, injection-molded cyclic olefin copolymer, top view; schematic)
fill window
[ INLET 20 µL ]==+==> A dried sphering diluent, 1:200 ======> [ RBC / PLT chamber, 20 µm deep ]
capillary |
metering +==> B dried RBC lysis + acridine orange, 1:10 ==> [ WBC chamber, 100 µm deep ]
|
+==> C dried sodium lauryl sulfate =========> [ Hb well, 0.13 mm path ] 540 / 880 nm
|
+==> D hemolysate + boronate conjugate =====> [ HbA1c membrane ] two-color reflectance
(Afinion principle) [ vent ]
side view: LED illumination: white, 470 nm, 540 nm, 880 nm
| | |
=========================== PMMA lid
--------------------------- chamber layer (20 µm and 100 µm spacers)
=========================== 0.17 mm optical floor
^ 20x / 0.45 objective, voice-coil focus, XY stage
^ Sony IMX585 camera, 500 nm long-pass + dual-band emission filter
Optics. A 20x/0.45 infinity-corrected objective and a Sony IMX585 sensor (3840 x 2160, 2.9 µm pixels) give 0.145 µm per pixel and a field of 0.557 x 0.313 mm (0.174 mm²) per image. An XY stage steps through fields; a voice coil keeps focus. White light gives brightfield; a 470 nm LED excites acridine orange, which stains DNA green and RNA and granules red.
Counting statistics. Counting error is set by how many cells you count (Poisson: CV = 1 / sqrt(N)). The chamber design is sized so that a normal sample produces enough cells inside the CLIA limits:
Chamber
Dilution
Depth
Cells per mm² at a normal value
Fields (13.9 mm²)
Cells counted
Counting CV
CLIA 2025 limit
RBC
1:200
20 µm
500 (RBC 5.0 million/µL)
80
about 6,960
1.2%
±4%
Platelets
same chamber
20 µm
25 (PLT 250,000/µL)
80
about 350
5.4%
±25%
WBC
1:10, RBCs lysed
100 µm
70 (WBC 7,000/µL)
80
about 970
3.2%
±10%
Scanning is adaptive: when platelets fall below 100,000/µL or white cells below 3,000/µL, the Station scans up to 200 fields so the count stays precise at the low values that matter clinically.
Settling. Cells must lie on the chamber floor to be in focus. From Stokes' law: sphered red cells settle 20 µm in about 15 seconds; white cells settle 100 µm in about 30 seconds; platelets settle 20 µm in about 2 to 3 minutes. The Station images red and white cells at 45 seconds and platelets at about 3 minutes, with a three-plane focus stack near the floor.
Output
Method
Precedent
CLIA 2025 limit
RBC, PLT, MCV, RDW
Brightfield imaging of sphered cells; size from diameter
OLO, HemoScreen
RBC ±4%, PLT ±25%
WBC and 5-part differential
Two-color acridine orange fluorescence plus brightfield; convolutional classifier
The 19 reported parameters: WBC, RBC, hemoglobin, hematocrit, MCV, MCH, MCHC, RDW, platelets, and neutrophils, lymphocytes, monocytes, eosinophils and basophils as both counts and percentages.
6. Engine 2: Chem Disc (centrifugal photometry)
Precedents: Abaxis Piccolo (100 µL, up to 14 analytes per disc, 12 minutes, 11 waived discs), Roche cobas b 101, biosurfit spinit, Samsung LABGEO PT10.
Chem Disc: plasma is separated, metered, diluted and shared among 27 cuvettes read by a ten-wavelength photometer.Plain-text version
CHEM DISC (80 mm diameter, three welded layers; radial layout, center at left, rim at right)
center ........................................................................................ rim
(spindle)
[ diluent blister 400 µL ] --pin--+
v
[ sample port ] --> [ sample chamber 120 µL ] --> [ separation arm ] --> [ RBC trap ]
| plasma (4,500 rpm for 75 s, about 680 g)
v
[ siphon valve ] --> [ meter 20 µL ] --> [ mix chamber, 1:20 ]
|
v
[ distribution ring: 24 reagent cuvettes + 2 reference + 1 interference index ]
|
xenon flash above (fiber-coupled) --- every cuvette crosses the beam once per turn
10 filtered photodiodes below: 340, 405, 450, 480, 500, 550, 570, 600, 660, 850 nm
Step
Time (s)
Speed
What happens
Load and pierce
0 to 10
0
A pin opens the diluent blister; disc warms to 37 °C
Prime
10 to 20
1,000 rpm
Diluent moves to its holding chamber
Separate
20 to 95
4,500 rpm
Cells pack into the trap (about 680 g at 30 mm radius)
Meter
95 to 110
600 rpm
Siphon primes; 20 µL of plasma is metered
Mix
110 to 140
±15° oscillation
Plasma and 380 µL of diluent mix
Distribute
140 to 160
3,000 rpm
About 13 µL fills each cuvette
React and read
160 to 720
1,500 rpm
Reagent beads dissolve; endpoint and kinetic reads every turn
Report
720
0
Results, interference indices, QC
Photometer. The baseline copies the proven architecture: a xenon flash lamp fired in sync with the rotor's optical index, a reference photodiode to normalize each flash, and ten interference-filtered photodiodes, each with a transimpedance amplifier and gated integrator feeding a 24-bit ADC. Averaging 100 flashes per reading targets absorbance noise below 0.0005. A cost-down path replaces the filter bank with a compact grating spectrometer for 400 to 850 nm plus a dedicated UV channel at 340 nm, where most enzyme assays read.
Metabolic + Lipid disc (17 measured analytes):
Analyte
Method
Read
CLIA 2025 limit
Sodium
Enzymatic, sodium-activated beta-galactosidase
405 nm kinetic
±4 mmol/L
Potassium
Enzymatic, potassium-activated pyruvate kinase with LDH
Interference indices. One cuvette holds diluted plasma with no reagent. Absorbance at 570 and 600 nm gives a hemolysis index, 450 and 480 nm an icterus index, and 660 and 850 nm a lipemia index. Each analyte has its own suppression threshold, set from CLSI EP07 interference studies. The Piccolo already reports these indices; the difference here is that the same indices also govern the other engines' results from the same draw.
Precedents: Boditech ichroma (about 60,000 readers), Wondfo Finecare, Getein 1100, ProciseDx (FDA-cleared time-resolved FRET, 2 to 5 minutes from fingerstick blood), Aidian QuikRead (CRP in 2 minutes).
Immuno strip and time-gated reader: europium keeps glowing after the short background fluorescence has died, so the reader waits, then measures.Plain-text version
IMMUNO STRIP (one of up to 6 in a pack), side view
[ sample port ] 15 µL heparin whole blood
v
[ plasma separation membrane ] -> [ europium nanoparticle conjugate ] -> [ nitrocellulose T line | C line ] -> [ wick ]
^
chase buffer blister (75 µL), released by a plunger 10 s after the sample
READER HEAD (moves across the strips on a lead screw, 0.1 mm steps)
365 nm LED pulse (20 µs) --> strip --> 615 nm band-pass filter --> SiPM --> gated integrator
time gate: wait 50 µs after the pulse, then integrate for 400 µs
europium emission lasts about 0.5 to 1 ms; plastic and plasma autofluorescence is gone in nanoseconds
Each strip reports one analyte against a lot-specific four-parameter logistic curve carried on the pack's NFC tag, normalized by the control line. Because the strip runs on whole blood, results are corrected for hematocrit using the Heme Card hematocrit from the same visit (QuikRead and the Afinion do similar corrections).
Analyte
Format
Pack
CLIA 2025 limit
Honest note
TSH
Sandwich
Pharmacy
±20% or 0.2 mIU/L
Lateral flow reaches second-generation sensitivity; monitoring of suppressed TSH goes to the hub
Free T4
Competitive
Pharmacy
±15% or 0.3 ng/dL
hs-CRP
Sandwich
Pharmacy
±30% or 1 mg/L
Ferritin
Sandwich
Pharmacy
±20%
25-OH vitamin D
Competitive
Pharmacy
not graded
Target: CDC Vitamin D Standardization-Certification Program criteria
PSA, total
Sandwich
Pharmacy
±20% or 0.2 ng/mL
Abnormal results confirmed at the hub; decisions belong to a clinician
hCG, quantitative
Sandwich
Pharmacy
±18% or 3 mIU/mL
Troponin I
Sandwich
Urgent care
±30% or 0.9 ng/mL
Contemporary, not high-sensitivity; an aid only; chest pain still means the emergency department
NT-proBNP
Sandwich
Urgent care
not in the CLIA table
Validated against a cleared predicate
D-dimer
Sandwich
Urgent care
not in the CLIA table
Exclusion use only with a validated cutoff
Procalcitonin
Sandwich
Urgent care
not in the CLIA table
CRP, wide range
Sandwich
Urgent care
±30% or 1 mg/L
Version 2 high-sensitivity module. High-sensitivity troponin and third-generation TSH need more signal than lateral flow gives. Two precedents show the way: magnetic nanoparticle immunoassays read by frustrated total internal reflection (Philips Minicare, now Siemens Atellica VTLi, 8 minutes) and spinning-chip immunoassays (SpinChip, 10 minutes). Both are patented, so this module needs a freedom-to-operate review or a license.
8. Engine 4: EC Card (urgent-care option)
Precedents: Abbott i-STAT CHEM8+ (CLIA-waived on venous blood), Siemens epoc, Nova StatSensor, Kalium Health.
A single-use card carries thin-film electrodes: ion-selective electrodes for sodium (calixarene ionophore), potassium (valinomycin), chloride and ionized calcium, a silver/silver chloride reference, enzyme electrodes for glucose (glucose oxidase), lactate (lactate oxidase) and creatinine (creatininase chain) read amperometrically at Prussian blue, a urease ammonium electrode for BUN, and a conductivity cell for hematocrit. Like the i-STAT, each card carries a calibrant pouch that wets the sensors before the sample arrives, which is what makes single-use electrochemistry precise. 65 µL, about 2 minutes. It adds ionized calcium and lactate to the menu and gives whole-blood electrolytes fast for urgent care.
Honest limit: a whole-blood potassium electrode cannot see hemolysis. The Station reports EC Card potassium only when the same draw's Chem Disc hemolysis index is below threshold, or when the operator records a free-flowing venous sample.
9. Pre-analytical quality, built in
Check
How
Action
Fill volume
Camera or optical sensor sees the meniscus at each inlet; the disc checks plasma volume
Consumable rejected before it runs
Sample age
Draw kit barcode timestamps collection
Glucose flagged after 30 minutes unspun; whole-blood consumables rejected after 60 minutes (targets, to be validated)
Clots and clumps
Imaging finds fibrin strands and platelet clumps; the disc sees plasma shortfall
Platelets suppressed, CBC flagged, recollect
Hemolysis
Disc index at 570 and 600 nm
Per-analyte suppression (potassium, AST, LDH and others)
Lipemia
Disc index at 660 and 850 nm
Per-analyte suppression
Icterus
Disc index at 450 and 480 nm
Per-analyte flag
Hematocrit extremes
Heme Card hematocrit; disc plasma yield
Immunoassay results corrected; flag below 25% or above 60%
Consumable integrity
NFC lot, expiry and temperature log; on-board control wells
Rejected
Operator
Badge login checked against training records
Untrained users locked out
Example reporting rule, as the patient sees it: "Potassium not reported: the sample shows hemolysis, which falsely raises potassium. A venous draw is recommended." No number is printed. Theranos diluted and reported; the Station suppresses and explains.
10. Electronics
Electronics: a compute module for imaging and software, a real-time controller for motors, optics and heat.Plain-text version
+------------------------------ MAIN COMPUTE -------------------------------+
| Raspberry Pi Compute Module 5 (8 GB) + Hailo-8L NPU on PCIe |
| Linux: UI, image pipeline, result engine, encryption, sync, audit log |
+-------+----------------+----------------+----------------+------------------+
| MIPI CSI-2 | SPI + UART | PCIe / NVMe | Ethernet, Wi-Fi, LTE
v v v v
[ IMX585 camera ] [ STM32H743 real-time controller ] [ 256 GB log ] [ network ]
| FreeRTOS, watchdog, interlocks
+-------------------+-----------+------------+-------------------+-------------------+
v v v v v
[ SPINDLE ] [ PHOTOMETER ] [ THERMAL ] [ IMMUNO SCANNER ] [ HEME STAGE ]
BLDC + FOC driver xenon flash, fiber, heater pads, TEC, 365 nm LED pulses, XY steppers,
optical index 10 filtered PDs, TIAs, 10k NTCs, 615 nm filter, voice-coil focus,
0-5,000 rpm gated integrators, 37.0 ± 0.1 °C SiPM, gated 4 LED channels
24-bit ADC PID loops integrator
+-------------------+------------------------+
v v v
[ EC FRONT END ] [ OPERATOR I/O ] [ POWER ]
AD5941 amperometry 10" touchscreen, 24 V in; 12 V, 5 V, 3.3 V rails;
and conductivity; 2D barcode, NFC lot supercapacitor for a safe stop;
LMP7721 + ADS1262 reader, status light lid lock solenoid
for the ISEs
Block
Example parts
Target cost at 1,000 units/yr
Compute
Raspberry Pi CM5 8 GB, Hailo-8L, carrier board
$180
Real-time control
STM32H743, motor and heater drivers, interlocks
$70
Spindle
BLDC motor, field-oriented-control driver, optical index
10-inch touchscreen, 2D barcode engine, NFC reader, status light
$140
Power and wiring
24 V supply, DC-DC converters, supercapacitor, harnesses
$70
Enclosure and mechanics
Sheet-metal frame, molded covers, drawers, lid lock, isolation mounts
$320
Assembly, test, calibration
Labor and fixtures
$250
Total
about $2,050
Pilot builds of 10 to 60 units will cost two to three times this.
11. Software and the cloud
Station OS. Embedded Linux on the Compute Module with secure boot, signed updates, encrypted local storage, and an offline mode that syncs later.
Real-time firmware. FreeRTOS on the STM32H743: motor control, flash timing, photodiode sampling, heater loops, the immunoassay scanner, the EC front end, interlocks and a hardware watchdog.
Pipelines. Chemistry: blank-corrected absorbance, rate or endpoint, lot calibration, interference rules, reportable-range checks. Hematology: focus stack, segmentation, classification, counts with confidence intervals, flags. Immunoassay: line profile, baseline fit, peak areas, test-to-control ratio, four-parameter curve. EC: sensor slopes from the calibrant, Nernstian conversion.
Result rules. Delta checks against the patient's previous results, critical values that trigger a documented call to the patient's clinician, and the suppression rules above.
Machine-learning governance. Models are locked in the field. Future model updates follow a predetermined change control plan agreed with FDA in advance.
Cloud. Fleet QC, lab director review, results routing with LOINC codes over HL7 v2 and FHIR R4, billing support (CPT codes with the QW modifier when waived), HIPAA, SOC 2, and the premarket cybersecurity requirements for connected devices, including a software bill of materials.
Quality loop: results that fail the rules are suppressed; a drifting analyte locks at that site until the lab director signs off.Plain-text version
[ Station run ] --> [ result engine ] --> range, interference, delta and critical-value rules --> [ pharmacist screen ]
| | |
| +--> not reportable? --> "not reported" + venous or hub option ------------+
v
[ QC record ] daily electronic QC | liquid QC per lot and every 30 days | on-board control wells
|
v (encrypted, signed)
[ THIMBLE CLOUD ] -------------------------------------------------------------------------------+
| Levey-Jennings per lot and site | Westgard multirule | patient moving averages (PBRTQC) |
| lab director review queue | critical-value call log | audit trail: who, what, when |
+-------------------------------------------------------------------------------------------------+
| |
v v
drift or QC failure --> that analyte locks at that site results --> HL7 v2 / FHIR R4 --> EHR, patient app
|
v
service ticket + lab director sign-off before unlock
Patient-based real-time QC (moving averages of, for example, sodium or MCV across all patients at one site) catches slow drift that liquid QC misses. Across a fleet, the cloud can compare every site to every other site running the same lot. This is the honest version of Theranos's remote-control software: it locks results out when quality fails instead of hiding failures.
12. Mechanical layout and scheduling
THIMBLE STATION, front view (target 42 W x 38 D x 32 H cm, under 15 kg)
+-------------------------------------------------------------------------------+
| [ 10" touchscreen ] ( status light ) |
| |
| +---------------+ +-----------------------------+ +--------------------+ |
| | BAY 1 | | BAY 2 | | BAY 3 | |
| | Heme Card | | Chem Disc (lid-locked) | | Immuno Pack | |
| | imaging | | 37 °C rotor | | up to 6 strips | |
| +---------------+ +-----------------------------+ +--------------------+ |
| |
| [ BAY 4: EC Card, urgent-care option ] [ NFC + barcode pad ] [ USB ] |
+-------------------------------------------------------------------------------+
imaging bay on elastomer isolation mounts; rotor on its own sub-frame
Chassis layout (target dimensions).
The scheduler runs the disc's high-speed separation spin during the Heme Card's 45-second settling period, so imaging never happens while the rotor is at 4,500 rpm. Imaging during the 1,500 rpm read phase is allowed because the bay is isolated and exposures are about 1 ms.
minute 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
collect (3 to 5 minutes before loading, outside the Station)
load #
Heme ######################### CBC + HbA1c, ~5 min
Disc ############################################################ 17 analytes, ~12 min
Immuno ########################################################################### up to 6 strips, 10-15 min
review ## pharmacist review
Run timing for a typical panel.
One rotor means 3 to 4 patients per hour. A busy urgent care needs two Stations or the version-2 dual-rotor chassis.
13. Consumables
Consumable
Storage target
Shelf-life target
COGS target at volume
Price to site (business plan)
Draw kit (lancet or arm collector, EDTA and heparin microtubes, wipes)
Room temperature
24 months
$1.50 (fingertip)
$3
Heme Card
Room temperature, foil pouch with desiccant
12 months
$3.50
$9
Chem Disc
2 to 8 °C, following the Piccolo precedent, until a room-stable bead formulation is validated
12 months
$5.00
$14 (Metabolic + Lipid), $12 (Extended)
Immuno strip
Room temperature
18 months
$1.50
$6
EC Card
Refrigerated with a room-temperature window, following the i-STAT precedent
6 months
$4.00
$10
14. Verification and validation plan
Question
Standard or method
Precision
CLSI EP05 (20 days, 2 runs, 2 replicates)
Linearity and reportable range
CLSI EP06
Interference (hemolysis, lipemia, icterus, common drugs)
CLSI EP07
Method comparison
CLSI EP09, at least 120 samples per analyte against a cleared laboratory analyzer, venous first
Detection limits
CLSI EP17
Reference intervals
CLSI EP28 verification, separately for capillary samples
Capillary versus venous
Paired study under IRB, at least 2,000 participants across ages, hematocrit ranges and skin types
CLIA waiver
Untrained operators at three or more intended-use sites; flex studies under FDA's waiver guidance
IEC 62304; cybersecurity under FD&C Act section 524B
Risk
ISO 14971
Acceptance: total error inside the CLIA 2025 limit for each analyte, with a sigma metric, (allowable error minus absolute bias) divided by CV, of 3 or more. Below that the analyte is not offered from capillary blood.
15. Honest limits
Capillary potassium is biased upward by hemolysis. The Station suppresses it when the hemolysis index is high and never "corrects" it.
Capillary total CO2 runs low when tubes are underfilled or exposed to air. It may end up venous-only.
Capillary platelets run lower than venous. Capillary CBCs get their own reference intervals.
No coagulation in version 1. PT/INR needs citrate at an exact ratio. A fingerstick INR strip is a proven add-on later (Roche CoaguChek is the precedent).
No blood gases. They need anaerobic handling.
Lateral flow is less precise than central-lab chemiluminescence. Third-generation TSH, high-sensitivity troponin, sex hormones, B12, folate, PTH, cortisol and serology go to the hub.
Throughput is about four patients an hour per Station.
Version-1 labeling excludes neonates, critically ill patients, patients on chemotherapy and patients with known blood disorders, as the Sysmex XW-100 label does.
Image classifiers can miss rare abnormal cells. Every flag sends images to the hub for human review.
Everything here is a target until verified. The business plan's stage gates say what is dropped if a target is missed.
New, and worth filing on where the prior art allows:
Four engines in one scheduled chassis, so a single visit returns a CBC, a metabolic and lipid panel and several immunoassays in about 15 minutes.
A Draw kit that splits the sample into dual-additive microtubes and timestamps it, linked to the Station by barcode.
Interference indices measured once and applied across engines: the disc's hemolysis index governs potassium on the EC card and results on the strips from the same draw.
CBC plus HbA1c on one imaging card from one EDTA application.
Fleet-wide patient-based QC across distributed point-of-care sites, with automatic per-analyte lockout and director sign-off.
A patient report that shows the day's QC status and lot for every result.
17. Roads not taken
Optical "scan the drop" (Raman, as Haemanthus proposes). Raman scattering is weak, blood fluoresces and hemoglobin absorbs strongly, and most analytes sit at micromolar to picomolar levels. It may measure glucose one day; it will not quantify 40 analytes soon.
One undifferentiated drop for 200 tests (Theranos). It violates both the anticoagulant rules and the volume budget.
A robot pipettor inside the box (the Edison). It was the Edison's mechanical failure point. Consumables that meter themselves remove it.
Silicon photonics first (Genalyte, SiPhox). Excellent multiplexing, but expensive chips and a slow regulatory path. A candidate for the version-2 immunoassay engine.
Digital microfluidics for everything (Baebies). Flexible and now cleared for PCR, but less proven for routine chemistry at volume. Worth watching.
The continuous patch (the 2003 vision). A different product. Continuous glucose monitors, the Libre Duo glucose-ketone sensor, Biolinq and Nutromics show where it is going; it is not the pharmacy box.